B. Braun Space is a modular infusion system that is smaller, lighter, and easier to handle than any other existing configuration. Leading the way into the future, this system meets all demands of an outstanding safety concept.
B Braun Syringe Pump Manual
. The initial training of the Infusomat® Space is to be performed by B. Braun sales and/or clinical personnel or other authorized persons. After each soft-ware update, the user must refer to the Instructions for Use to review changes to the device and software. Caution: Ensure the unit is properly positioned and secured (see ch.
B. Braun Space can also be integrated into the data communications network of every advanced hospital operation.
Performance data
Automatic rate calculation by calculating volume over time
Dose rate mode:
Automatic calculation of delivery rate based on the entry of drug concentration in conjunction with the desired dose rate (e.g.: mg/kg/min)
Drug Library:
- Up to1500 drug names including therapy data and information can be stored in30 categories
- Up to 50 care units and15 patient profiles possible
- Drug specific soft and hard limits as well as default values can be specified
Operator interface
- Standardized and intuitive user interface for all B. Braun Space infusion pumps
- Due to menu operation and arrow keys the user has to look at the display when doing adjustments
Safety concept
- Integrated piston brake prevents free-flow during syringe change
- Automatic drive technology ensures startup and convenience when changing syringes
- Automatic bolus reduction triggered by occlusion alarm
- Data lock on3 safety levels (parameters and disposables lockable)
- LED alarm indicator with clear alarm message in display
Related products
- Parts
- Videos
- News
DESCRIPTION
It lets you plan for the future today.The Outlook Safety Infusion System helps to ensure that the right patient receives the right medication in the right dose from an authorized clinician at point-of-care. The Outlook 100 incorporates dose limits to help reduce human errors in medication administration.
FORUMSView All (30)
Ask a New Question1Reply6 months ago | 6 months agoElectrical schematics Does anyone know where to find the schematic for this device. TiaReply |
-ivpump.com 9 months ago | 9 months agoGo-No/Go Gauge Does anyone know where I can purchase the Go-No/Go Gauge or the dimensions of the gauge for the Yoke Adjustment Test?Reply |
0Replies2 years ago | 2 years agoPM notes to keep in mind Batteries in this unit are NiCD alarm batteries which I found always are dead. The lithium memory batteries last forever evidently (9 years+), and main batteries a couple of years or so depending if the nursing staff will plug them in. The petal covers wear thin and have to be replaced annually for me (every other PM cycle). I check the gaps with the gap tool and found not likely to be off, but best to check the gap.I hate changing the alarm batteries and grimace when I pull one out because it has nine leads that all have to be unsoldered on tiny pads through the board. Why did they use NiCD batteries for goodness sakes??The green boards are easier than the blue ones to PM and change batteries with.Reply |
DOCUMENTS / MANUALSView All
FEATURES
B Braun Infusion Pump User Manual Instructions
- Promotes patient safety by helping to avoid accidental under- or overdosing using audible and visual alerts
- Provides flexibility for each care area—ability to customize dose limits to your hospital drug formulary
- Promotes clinician choice with dose limit override when limits are exceeded during manual programming and titration
- Visually indicates dose selection with DoseGuard software
- Easy to learn
- Manual operation RATE, VOLUME, RUN
- Automatic Dose Rate Calculator
- Large, highly visible keys
- Reduce nuisance air-in-line alarms with air-trapping cassette
- Stack-able modular design reduces bedside cord clutter and errors due to channel confusion
- Comprehensive line of needle-free administration sets including UV light resistant tubing that protects drug efficacy
B Braun Infusion Pump Manual
SPECIFICATIONS
Flow Accuracy | 5 % |
Flow Rate Range | 0.1 to 999.9 ml/h |
Infusion Pump Type | General-Purpose |
KVO Rate | 3 ml/h or selected rate whichever is lower |
Volume to be Infused | 0.1 to 999.9 ml |
Height | 4.5 in |
Length | 8.75 in |
Weight | 5 kg |
Width | 12 in |
B Braun Infusion Pump User Manual Free
Additional Specifications
Braun Infusion Pumps User Manual
Power: 120 VAC