3m Bair Hugger 775 User Manual

  1. 3m Bair Hugger 775 Manual
  2. 3m Bair Hugger 775
  3. Bair Hugger 775 Keypad
  4. 3m Bair Hugger 775 Manual

Bair Hugger 500-OR

User manual

Refer to the 3M™ Bair Hugger™ Model 775 warming unit Operator’s Manual for a complete list of instructions for use, preventative maintenance, warnings and cautions. For more information about the Bair Hugger warming system, please contact your 3M representative, call us at 3957-228-800-1 or visit www.bairhugger.com. Bair Hugger Model 505 and Model 500/OR Temperature Management Units Operator’s Manual. Control Panel Features of the Model 500/OR Temperature Management Unit. OVER HEAT INDICATOR. The Over Heat Indicator illuminates and an audible alarm sounds when an over-temperature condition is detected.

680 KB Download prohibited by Arizant (3M)

Bair Hugger 505

User manual
1.2 MB Download prohibited by Arizant (3M)

Bair Hugger 505

User manual
3.1 MB Download prohibited by Arizant (3M)

3m Bair Hugger 775 Manual


Bair Hugger 750

User manual
1.2 MB Download prohibited by Arizant (3M)

Bair Hugger 775

User manual
2.1 MB Download prohibited by Arizant (3M)

Bair Paws 850

User manual
950 KB Download prohibited by Arizant (3M)

Bair Paws 875

User manual775
1.1 MB Download prohibited by Arizant (3M)

Cincinnati Blanketrol 2

User manual
2.0 MB Download

CSZ Hemotherm 400MR

User and service manual
2.7 MB Download

Smiths EQ-5000

User manual3m bair hugger 775 manual
960 KB Download prohibited by Smiths Medical. Support is not desired.
TSCI Mistral-Air Plus MA-1100
User manualBair
1.9 MB DownloadWith kindly support of 'The 37Company'
One moment please...
  • Parts
  • Videos
  • News
DESCRIPTION
The future gold standard in patient warming.

Arizant's latest advancement in forced-air warming - the Bair Hugger Model 775 temperature management unit - is the future gold standard in patient warming.

FORUMSView All (4)
Ask a New Question

3m Bair Hugger 775

1

Reply
-JStilley
3 years ago
3 years agoPM Procedure For The 3M Forced-Air Arizant Bair Hugger model 700
The PM for the Bair Hugger is fairly simple. As with most PM's you will need to do a currant leakage check with a standard Safety Analyzer. Assure the ohms are below .5 and uA is below 300. Next you will want to turn on the device and allow it to heat up. You will be testing three different temperature ranges: Low 32°C (±1.5), Med 38°C (±1.5), High 43°C (±1.5) using a temperature probe that is fitted into tubing that fits onto the Bair Hugger hose port. You may test the overheating alarm by occluding the hose until it overheats and alarms. Finally you will need to replace the filter located in the base of the Bair Hugger. This is a high efficiency 0.2¼m filter, part # 90047 from 3M. This filter will need to be changed annually. This concludes the PM Procedure for the 700 series Bair Hugger. If I missed anything or you need further details or information please contact me. Thank you!Reply
-JStilley
3 years ago
3 years agoPM Procedure For The Forced-Air Arizant Bair Hugger
The PM for the Bair Hugger is fairly simple. As with most PM's you will need to do a currant leakage check with a standard Safety Analyzer. Assure the ohms are below .5 and uA is below 300. Next you will want to turn on the device and allow it to heat up. You will be testing three different temperature ranges: Low 32°C (±1.5), Med 38°C (±1.5), High 43°C (±1.5) using a temperature probe that is fitted into tubing that fits onto the Bair Hugger hose port. You may test the overheating alarm by occluding Reply

1

Reply
-Putman
7 years ago
7 years agoNeed Quote

Is MedWrench an authorized dealer for Arizant? Are you a small business? Are you registered as a federal contractor?

Reply
DOCUMENTS / MANUALSView All

Bair Hugger 775 Keypad


FEATURES

3m Bair Hugger 775 Manual

  • Adjustable airflow, precise temperature delivery and ease of use from a single, quiet warming unit
  • Two airflow settings and hose-end temperature sensing ensure precise temperature delivery and compatibility with all Bair Hugger blanket styles - pediatric to adult
  • Three temperature sensors and adjustable airflow provide safe, quick temperature response and greater heat transfer
  • Tracks and reports forced-air warming therapy duration, over-temperature information, calibration, built-in hour meter and fault codes all via the front panel - no need to open the unit
  • Compatible with Bair Hugger 241® blood/fluid warming set